Set Gage Zero
This command sets the zero-load value on a strain gage channel. In effect, the Set Gage Zero
command tells the 518 that there is presently no load on the channel load cell. Typically, this
command is issued just prior to a Set Gage Span command. Together, the Set Gage Zero and Set
Gage Span commands may be used to calibrate a strain gage channel.
COMMAND: (176 + CHAN)
RESPONSE: NONE
See the Strain Gages section for more details on strain gage commands.