Sensoray Model 618/619 Smart A/D 19 Specifications
9 Specifications
9.1 General Specifications
Table 6: General Specifications
Category Parameter Condition Specification
Power Quiescent +5VDC, ±5% 100mA, typical.
+12VDC, ±5% 35mA, typical.
-12VDC, ±5% 35mA, typical.
Temperature Operating range Guaranteed accuracy -25°C to +85°C
Sense inputs CMRR CMV < 5V, f < 1KHz 80dB, minimum
CMV Guaranteed accuracy ±5.0V, maximum
Input impedance SCC disabled 1000 M, minimum
SCC enabled 1 M, typical
Input protection Continuous ±20V CMV, maximum
2 seconds, max. ±70V CMV, maximum
A/D
converter
Type Integrating, 16-bit resolution.
Measurement period,
per channel
60Hz reject mode
(default mode)
16.7 milliseconds, typical (default mode).
4.0 milliseconds, typical (high-speed mode).
50Hz reject mode 20 milliseconds, typical (default mode).
4.8 milliseconds, typical (high-speed mode).
Total time slot,
per channel
60Hz reject mode
(default mode)
22 milliseconds, maximum (default mode).
9 milliseconds, maximum (high-speed mode).
50Hz reject mode 25.3 milliseconds, maximum (default mode).
11 milliseconds, maximum (high-speed mode).
Excitation
source
Pulse width Applied for duration of channel time slot.
Type Strain/pressure gage 10VDC, 40mA maximum.
400/3K ranges 1.2mADC.
600K range 5VDC in series with 4K.
Bus interface Type PCI Slave, 32-bit, 33MHz
Address
requirements
Three I/O address blocks: sizes 32, 128 and 256 bytes.
PCI configuration registers require an additional amount.